BackgroundAlthough the modular sleeve concept for femoral implants has several advantages, studies reporting on greater than 20-year outcomes are limited. This study aimed to report the mean 23-year results of total hip arthroplasty (THA) using a modular stem with metaphyseal fixation sleeve. MethodsThis study reviewed primary THAs performed using a modular stem between October 1994 and April 1998. Of the 272 hips (221 patients) identified, 220 hips (177 patients) satisfying the minimum 20-year follow-up were included. The Harris Hip Score was used for clinical evaluation. Final hip radiographs were analyzed to evaluate the implant stability and the extent of femoral osteolysis. The mean follow-up duration was 23 years (range, 20 to 28). ResultsThe mean Harris Hip Score improved from 38 points (range, 26 to 67) preoperatively to 90 points (range, 71 to 100) at the final follow-up (P < .001). Femoral osteolysis was observed in 146 hips (67.1%), but 145 were located proximal to the osseo-integrated sleeve. Overall, 4 femoral revisions were performed; 2 for chronic infection, one for aseptic loosening, and one for periprosthetic femoral fracture. Of the 19 (8.6%) periprosthetic femoral fractures, the most common type was Vancouver AG (13 hips). One (0.5%) asymptomatic distal stem fracture occurred, while no complications were identified at the stem-sleeve junction. Implant survivorship free of any femoral revision was 98.3% at 23 years. ConclusionsA THA using a modular stem with metaphyseal fixation sleeve demonstrated 98.3% stem survivorship with excellent clinical outcomes at a mean follow-up of 23 years in non-obese patients.