This study was designed to analyze the incidence of "dual tachycardia"-ventricular tachycardia (VT) or ventricular fibrillation (VF) preceded by paroxysmal atrial tachycardia (AT) or atrial fibrillation (AF)-in patients receiving dual-chamber implantable cardioverter defibrillators (ICDs). Paroxysmal AT/AF occurs commonly in patients who receive ICDs for the treatment of life-threatening VT/VF. Although AF is associated with an adverse prognosis in the setting of structural heart disease, the relationship between AT/AF and VT/VF is unclear. We followed 537 patients undergoing implantation of the Jewel AF ICD (Model 7250, Medtronic, Minneapolis, Minnesota) for 11.4 +/- 8.2 months. These included 398 patients with a history of at least two episodes of AT or AF during the preceding year as well as 139 patients enrolled because of VT/VF alone. There were 233 dual tachycardia episodes in 45 patients during follow-up. Overall, 8.9% of episodes detected as VT/VF were dual tachycardias, and 20.3% of patients with VT/VF had at least one dual tachycardia episode. The median duration of AT/AF preceding the first VT/VF detection was 1.09 h (25% to 75% quartile 0.24 to 33.4 h). When AT/AF continued between two consecutive VT/VF detections, the median interdetection interval was 11 min. When AT/AF terminated either because of a ventricular therapy or spontaneously, the median interdetection interval was prolonged to 71 h (p < 0.001). Dual tachycardia is common in ICD recipients with a history of AT/AF. The duration of AT/AF preceding the first VT/VF detection is < or =1 h about 50% of the time. Termination of the AT/AF significantly delays the time to the next VT/VF detection.