Transferrin, a major glycoprotein has different isoforms depending on the number of sialic acid residues present on its oligosaccharide chain. Genetic variants of transferrin as well as the primary (CDG) & secondary glycosylation defects lead to an altered transferrin pattern. Isoform analysis methods are based on charge/mass variations. We aimed to compare the performance of commercially available capillary electrophoresis CDT kit for diagnosing congenital disorders of glycosylation with our in-house optimized HPLC method for transferrin isoform analysis. The isoform pattern of 30 healthy controls & 50 CDG-suspected patients was determined by CE using a Carbohydrate-Deficient Transferrin kit. The results were compared with in-house HPLC-based assay for transferrin isoforms. Transferrin isoform pattern for healthy individuals showed a predominant tetrasialo transferrin fraction followed by pentasialo, trisialo, and disialotransferrin. Two of 50 CDG-suspected patients showed the presence of asialylated isoforms. The results were comparable with isoform pattern obtained by HPLC. The commercial controls showed a <20% CV for each isoform. Bland Altman plot showed the difference plot to be within +1.96 with no systemic bias in the test results by HPLC & CE. The CE method is rapid, reproducible and comparable with HPLC and can be used for screening Glycosylation defects.