BackgroundThis diagnostic prospective study compared the feasibility and diagnostic accuracy of Pocket-size Ultrasound Devices (PUDs) against standard ultrasound (US) in detecting liver steatosis using the controlled attenuation parameter (CAP) and liver biopsy as reference standards. Materials and MethodsConsecutive patients with chronic liver diseases were assessed for the presence of steatosis using PUD and US. A CAP cut-off value >275 dB/m was applied to establish ≥S1. A 26-patient subgroup underwent liver biopsy. PUD reproducibility was evaluated using Cohen's k statistic. Diagnostic accuracy of PUD and US was given as Sensibility (Sn), Specificity (Sp), Positive and Negative Predictive Values (PPV, NPV), positive and negative Likelihood Ratio (LR+, LR-). Results81 consecutive patients (69% males) with multiple etiologies were enroled. PUD inter-observer agreement was good (k 0.77, 95%CI 0.62–0.93). PUD and US identified ≥S1 according to CAP values respectively with Sn 0.87, Sp 0.61, PPV 0.49, NPV 0.91, LR+ 2.04, LR- 0.07, AUROC 0.74 and Sn 0.96, Sp 0.54, PPV 0.47, NPV 0.97, LR+ 2.10, LR- 0.07, AUROC 0.75. ConclusionsPUD shows good reproducibility and diagnostic accuracy in ruling liver steatosis out, representing a useful point-of-care tool to avail of hepatologists interested in excluding NAFLD, but with basic US skills.