Background and Objectives: This study's objective was to investigate the influence of increased scan speed and pitch on image quality and nodule volumetry in patients who underwent ultra-low-dose chest computed tomography (CT). Material and Methods: One hundred and two patients who had lung nodules were included in this study. Standard-speed, standard-pitch (SSSP) ultra-low-dose CT and high-speed, high-pitch (HSHP) ultra-low-dose CT were obtained for all patients. Image noise was measured as the standard deviation of attenuation. One hundred and sixty-three nodules were identified and classified according to location, volume, and nodule type. Volume measurement of detected pulmonary nodules was compared according to nodule location, volume, and nodule type. Motion artifacts at the right middle lobe, the lingular segment, and both lower lobes near the lung bases were evaluated. Subjective image quality analysis was also performed. Results: The HSHP CT scan demonstrated decreased motion artifacts at the left upper lobe lingular segment and left lower lobe compared to the SSSP CT scan (p < 0.001). The image noise was higher and the radiation dose was lower in the HSHP scan (p < 0.001). According to the nodule type, the absolute relative volume difference was significantly higher in ground glass opacity nodules compared with those of part-solid and solid nodules (p < 0.001). Conclusion: Our study results suggest that HSHP ultra-low-dose chest CT scans provide decreased motion artifacts and lower radiation doses compared to SSSP ultra-low-dose chest CT. However, lung nodule volumetry should be performed with caution for ground glass opacity nodules.