The objective of this study was to compare the resolution and density appropriate to diagnosis in chest PA radiography. In comparing the resolution, we radiographed with conventional radiography, computed radiography(CR) and digital radiography(DR) using the linear resolution phantom(Nuclear Associates-Carle Place. N.Y.). 2 radiologists and 3 radiological technologists read the resolution value by the blind test. DR, conventional radiography and CR measured 3.95, 3.58, 3.48 resolution value respectively. In analysing the density, we chose the fifty normal chest CR and DR and conventional film. We estimated the density using by densitometer(X-rite company-Model 301) in seven regions(lung field, lung field margine, mediastinum I, mediastinum II, heart shadow I, heart shadow II, diaphragm) of chest film. We adapted to analysis the Japanese chest X-ray evaluating method and table. It was scored 0(farthest density value) to 2(nearest density value). DR scored 2 at mediastinum I, mediastinum II, heart shadow I, heart shadow II and diaphragm. On the contrary with, CR scored 2 at lung field and lung field margine. Consequently, DR superior than CR and conventional radiography film compairing density and resolution. It was due to small pixel size and post processing algorithm with digital radiography.