Improved understanding is required of how the mechanics of the fall affect hip fracture risk. We used a hip impact simulator to determine how peak stresses at the femoral neck were affected by pelvis impact angle, hip abductor muscle force, and use of a wearable hip protector.We simulated falls from standing (2 m/s impact velocity) involving initial hip abductor muscle forces of 700 or 300 N. Trials were acquired for impact to the lateral aspect of the greater trochanter, and impact to the pelvis rotated 5°, 10° and 15° anteriorly (positive) or posteriorly (negative). Measures were acquired with and without a commercially available hip protector. During trials, we measured three-dimensional forces with a load cell at the femoral neck, and derived peak compressive and tensile stresses.Peak compressive stress increased 37% (5.91 versus 4.31 MPa; p < 0.0005) and peak tensile stress increased 209% (2.31 versus 0.75 MPa; p < 0.0005) when the pelvis impact angle changed from 15° anterior to −15° posterior. For lateral impacts, the peak tensile and compressive stresses averaged 73% and 8% lower, respectively, in the 700 N than 300 N muscle force condition, but the effect was reversed for anteriolateral or posteriolateral impacts. The attenuation in peak compressive stress from the hip protector was greatest for posteriolateral impacts (−15 to −5°; 36–41%), and least for anteriolateral (+15°; 10%).These results clarify the effects on hip fracture risk during a fall of pelvis impact angle and muscle forces, and should inform the design of improved hip protectors.