The determination and maintenance of constant low but non-zero levels of oxygen is critical in the study of the radiation chemical interactions of nitroimidazoles in mammalian cells in vitro. As well, many of these chemicals have increased toxicity toward hypoxic compared to aerobic cells, although absolute hypoxia probably is not required. Both of these phenomena must be investigated in systems where significant consumption of oxygen takes place, either through radiation depletion or by cellular metabolism. In this paper an analysis has been made of the form of oxygen depletion in stirred cell suspensions with overlying gas phase, and it has been found to conform to the relationship (C(t) - C/sub infinity/) = (C(0) - C/sub infinity/) exp(-k/sub 1/t), where C/sub infinity/ = C/sub g/ - R/k/sub 1/. Here C(t) is the oxygen tension throughout the solution; C/sub g/, the equivalent level in the overlying gas phase; R (concentration units per sec), the depletion rate; k/sub 1/ (sec/sup -1/), a physical constant independent of oxygen concentration and depletion rate; and C/sub infinity/, the oxygen level in solution approached at long times. This relationship has been confirmed in detail using a Clark-type oxygen sensor and a high-stability amplifier design due to Koch.more » Since oxygen levels down to a few hundred parts per million can be determined with accuracy, it has been possible to measure precisely the oxygen levels present in our experimental systems. Implications of these results for the interpretation of data obtained in stirred cell suspension with overlying gas phase under conditions of consumption are discussed.« less