Abstract Introduction/Objective The management of internal quality control (IQC) in Sidra Medicine Clinical Chemistry Division has been evaluated in order to promote a more consolidated and efficient process of IQC management. The statistical data produced from Cerner QC Module are transferred to IQC review templates consisting of formulas to auto- calculate parameters such as multiple of expected QC failure frequency and desirable comparison limit between analyzers. The IQC review and documentation process using the in-house excel template requires several hours to complete, hence a faster and more efficient IQC management module is required. The main objective of this study is to improve the initial IQC management set up, work flow and review procedures and to implement Biorad Unity Real Time (URT) program to develop a more efficient IQC management system. Methods The URT software has been recently configured and implemented to consolidate and streamline IQC management. URT is built through Sidra Medicine IT Enterprise level which allows multiple users to login. IQC data are downloaded using scripts from Cerner which are filtered through Biorad Unity Connect (UC) software. Additional quality tools are also explored such as various user defined statistical reports, IQC analysis using peer reviewed total allowable error (TeA) and assignment of the most appropriate Westgard rules. Determination of sigma metrics and uncertainty of measurement is also performed using the URT application. Results The generation of any IQC report is less cumbersome and time consuming as compared with the previous process. However, some user defined formulas in the IQC templates are not found on the URT reports. The URT Levey Jennings chart are also more user friendly and directly compares the daily IQC data with Unity inter-laboratory peers enabling the production of instant and monthly reports through QCNet site when assay investigation is required and for IQC report documentation. Conclusion The combination of Cerner IQC, Unity Real-time, QCNet Inter-laboratory reports and in house IQC templates produce a high level and very detailed IQC review which effectively evaluate assay performance to assist on IQC troubleshooting and root cause analysis to be able to apply the most appropriate corrective actions.
Read full abstract