To assess the impact of inserting an ophthalmic viscoelastic device (OVD) into the dexamethasone (DEX) implant needle on pellet velocity in simulated vitrectomized eyes. DEX implants were injected into a calibrated ex vivo test chamber filled with balanced salt solution (BSS). All DEX implants were administered by the same physician and aiming for the same button depression time. In Group 1, three DEX implant injections were performed without an OVD, while in Group 2, the OVD was safely inserted into the DEX implant needle using a 27-gauge cannula just prior to injection. The slow-motion video mode of the IPhone 14 was utilized to record the procedures and calculate time and distance measurements. Group 1 exhibited a mean velocity of 450 mm/sec for the DEX pellet in BSS, compared to 54.57 mm/sec in Group 2. Furthermore, DEX pellets in Group 1 had horizontal displacements of 24, 29, and 31 mm, while those in Group 2 had displacements of 17, 16, and 15 mm. Injecting OVD into the DEX implant needle significantly decreases the velocity of the free pellet in BSS by 87.87%. This modification may help prevent potential velocity-related complications linked to DEX implantation in vitrectomized eyes.