Extracorporeal centrifugal blood pumps are used to treat cardiogenic shock. Owing to the imbalanced excitation or initial assembly configurations, the variation in the impeller axial position has the potential to affect the blood pump performance. This study compared the hydrodynamics and hemolysis outcomes at different impeller axial positions via numerical simulations. The result shows that pressure difference of the blood pump decreased with increasing impeller axial position, with decreasing by 4.5% at a flow rate of 2 L/min. Under axial impeller motion close to the top pump casing, average wall shear stress and scalar shear stress reached their maximum values (64.2 and 29.1 Pa, respectively). The residence time in the impeller center hole and bottom clearance were extended to 0.5 s by increasing impeller axial position. Compared to the baseline blood pump, hemolysis index increased by 12.3% and 24.3% when impeller axial position is 2.5 and 4.0 mm, respectively. As a novelty, the findings reveal that the impeller axial position adversely affects hemolysis performance when the impeller is close to the pump casing. Therefore, in the development process of centrifugal blood pumps, the optimal axial position of the impeller must be defined to ensure hemodynamic performance.