Coronary computed tomography (CT) angiography has become amajor cornerstone in the diagnostic workup of cardiologic patients, particularly for evaluation of the coronary arteries and preprocedural planning of interventions for structural heart disease. Despite the possible problems that intensive electromagnetic radiation (including X‑rays) might cause when directly impacting on implanted cardiac devices, cardiac CT is asafe diagnostic test and should not be withheld from patients with devices if properly indicated. Sufficient image quality is paramount for the evaluation; hence, special attention should be paid to alow heart rate (< 60 bpm) and sufficient compliance with breathing instructions. Furthermore, pacemaker or implantable cardioverter-defibrillator (ICD) leads may cause metal artifacts, especially around the lead tip. Their dense material causes beam hardening and streak artifacts which may result in reduced image quality and limited diagnostic assessability. The prevalence of such artifacts depends not only on lead material but also on lead positioning relative to the gantry plane. Metal artifacts are more frequent in patients with unipolar leads and shock coils, which can impair the assessment of coronary arteries, mainly of the right coronary artery (RCA). Artifacts caused by left ventricular (LV) leads of cardiac resynchronization therapy (CRT) systems tend to affect assessment of the left circumflex artery (LCX). By using dual energy CT and postprocessing algorithms, the impact of artifacts can be reduced and diagnostic image quality can be achieved in most cases. Unfortunately, the actual occurrence of such artifacts or the degree of impairment of image quality cannot be reliably predicted.