Various catheter designs are appearing for Pulsed Field Ablation (PFA). It is unclear if they differ in terms of safety and efficiency. PFA studies have reported hemolysis, kidney injury, high troponin, among other side effects. Using a CT-derived computer model, we compared catheter designs using two metrics: (1) efficiency: power delivered to an atrial wall target, expressed as a percent of total generator power; and (2) safety: electric current to achieve 90% transmurality (since more energy causes more collateral effects), as well as the corresponding electrode current density (ECD), a heat and bubble metric. The following catheter designs were compared: penta-spline basket, Nitinol spheres (focal 9 mm and large 1-shot), circular, balloon, and flex-circuit. Target was a 6 × 47 mm circumferential segment of atrial wall at LPV antrum. Transmurality was defined as percent of target having >600 volts per centimeter (V/cm) electric field needed for electroporation. Efficiency was 0.9, 1.4, 2.7, 5.9, 10, and 12% for the large 1-shot and 9 mm Nitinol spheres, penta-spline, circular, flex spline, and balloon catheters, respectively. Regarding safety, currents for 90% transmurality were 70, 39,36,12.5, 5.3, and 4 Amps for the same respective catheters, with less being safer. ECD was 124, 25, 74, 83, 41, and 31 A/cm2, respectively. Computer models demonstrated a remarkable range in efficiency among catheters studied. Those having less atrial blood exposure had the highest efficiencies, with factors of up to 13X more efficiency compared to exposed ones. Higher efficiency designs have less collateral current and are safer. Confirmatory in-vivo studies are required.