PurposeTo extend the knowledge-based (KB) automatic planning approach to CyberKnife in the case of Stereotactic Body Radiation Therapy (SBRT) for prostate cancer. MethodsSeventy-two clinical plans of patients treated according to the RTOG0938 protocol (36.25 Gy/5fr) with CyberKnife were exported from the CyberKnife system to Eclipse to train a KB-model using the Rapid Plan tool. The KB approach provided dose-volume objectives for specific OARs only and not PTV. Bladder, rectum and femoral heads were considered in the model. The KB-model was successfully trained on 51 plans and then validated on 20 new patients. A KB-based template was tuned in the Precision system for both sequential optimization (SO) and VOLO optimization algorithms. Plans of the validation group were re-optimized (KB-TP) using both algorithms without any operator intervention and compared against the original plans (TP) in terms of OARs/PTV dose-volume parameters. Paired Wilcoxon signed-rank tests were performed to assess statistically significant differences (p < 0.05). ResultsRegarding SO, automatic KB-TP plans were generally better than or equivalent to TP plans. PTVs V95% was slightly worse while OARs sparing for KB-TP was significantly improved. Regarding VOLO optimization, the PTVs coverage was significantly better for KB-TP while there was a limited worsening in the rectum. A significant improvement was observed in the bladder in the range of low-intermediate doses. ConclusionsAn extension of the KB optimization approach to the CyberKnife system has been successfully developed and validated in the case of SBRT prostate cancer.