BackgroundProximal humerus fractures are often treated with a fixed-angle titanium plate osteosynthesis. Recently, plates made of alternative materials such as carbon fibre-reinforced polyetheretherketone (CFR-PEEK) have been introduced. This study presents the postoperative results of patients treated with a CFR-PEEK plate.MethodsPatients with proximal humerus fractures treated with a CFR-PEEK plate (PEEKPower™ Humeral Fracture Plate (HFP)) were included. In follow-up examination, age and gender adjusted Constant-Murley Score (ACS), Subjective Shoulder Value (SSV), Quick Disabilities of the Arm, Shoulder and Hand Score (QDASH) and pain score (Visual Analog Scale (VAS)) were analyzed. General condition at follow-up was measured by European Quality of Life 5 Dimensions 3 Level Version (EQ-5D-3L). Range of motion was recorded. In addition, radiographs at follow-up, unfavorable events and revision rate were analyzed.ResultsIn total, 98 patients (66.0 ± 13.2 years, 74 females, 24 males) were reexamined. Mean follow-up was 27.6 ± 13.2 months. There were 15 2-part, 28 3-part and 55 4-part fractures. The functional scores showed good results: SSV 83.3 ± 15.6%, QDASH 13.1 ± 17.0 and ACS 80.4 ± 16.0. A 4-part-fracture, head split component, nonanatomic head shaft reposition and preoperative radiological signs of osteoarthritis were significant negative predictors for poorer clinical scores. Unfavourable events were observed in 27 patients (27.6%). Revision surgery was performed in 8 (8.2%) patients. Risk factors for an unfavourable event were female gender, age of 50 years and older, diabetes, affected dominant hand, 4-part fracture, head split and preoperative radiological signs of osteoarthritis.ConclusionThere are several advantages of the CFR-PEEK plate (PEEKPower™ Humeral Fracture Plate (HFP)) such as the polyaxial screw placement and higher stability of locking screws. In summary, the CFR-PEEK plate osteosynthesis is a good alternative with comparable clinical results and some biomechanical advantages. Proximal humerus fractures show good clinical results after treatment with a CFR-PEEK plate. The revision rate and the risk of unfavorable events are not increased compared to conventional titanium plate osteosynthesis.Level of evidenceIV
Read full abstract