Standard hydrodynamic tests of the new prosthetic valve have been carried out for comparing with the other existing valve designs. A new method for the heart valve prosthesis testing in a tornado-like flow has been developed. The valve function has been verified in a swine excluding the anticoagulation treatment during the period of time exceeding six months. Results The significant advantage of the new prosthesis in the standard hydrodynamic tests has been demonstrated. The tests in the tornado-like flow have shown that only this prosthesis allows maintaining the pattern, the head and flow rate characteristics of the tornado-like jet. Upon implanting the new prosthesis in the aortic position in a swine, the good performance of the valve without anticoagulation therapy has been confirmed in the course of more than six months. Conclusion Obtained has been the evidence of the merits of the new mechanical aortic valve owing to the due consideration of the hydrodynamic peculiarities of the aortic blood flow and the creation of the design providing the proper hydrodynamic compatibility.