We present a measurement setup consisting of two fluid-filled pressure chambers to mimic the mechanical stress likely to that of small body movements on biomedical flexible micro-electrode arrays for the analysis of various degradation mechanisms. Our main goal was the simulation of micro-motions in fluid conditions, while maintaining an electric access to the device. These micro-motions would be likely to those occurring in the human body caused by the intracranial pressure in magnitudes of 7-25 mmHg, which translates to a fluid pressure of 9-33 mbar. Furthermore, severe mechanical stress can be administered to the samples under the previously mentioned environment. Therefore, a flexible, polyimide-based sample with various metal test structures was fabricated and analyzed in the presented measurement setup. A comparison of the elongation of the sample's surface as a function of the applied hydrostatic pressure is given with computer simulations.