The research has developed an enzyme biosensor for the detection organophosphate pesticide residues. The biosensor consists of a pair of screen-printed carbon electrode (SPCEs). One of electrodes contains immobilized organophosphate hydrolase (OPH) on a chitosan membrane by cross-linking it with glutaraldehyde. The area of the electrodes was optimized to 3, 5, and 7 mm2. The OPH was isolated from Pseudomonas putida, and was purified by the ammonium sulfate precipitation method, with 6444 ppm (A) and 7865 ppm (B). The organophosphate pesticide samples were 0–100 ppb in tris-acetate buffer 0.05 M, pH 8.5. The results showed that the best performance of the biosensor was achieved by the enzyme A with an electrode area of 5 mm2. The sensitivity of the biosensor was between 3 and 32 µS/ppb, and the detection limit for the organophosphate pesticides was 40 ppb (diazinon), 30 ppb (malathion), 20 ppb (chlorpyrifos), and 40 ppm (profenofos).