PurposeTo evaluate the long-term efficacy and rotational stability of the AcrySof toric intraocular lens (IOL) in correcting preoperative astigmatism in cataract patients.MethodsThis prospective observational study included 30 eyes from 24 consecutive patients who underwent implantation of an AcrySof toric IOL with micro-coaxial cataract surgery between May 2008 and September 2008. Outcomes of visual acuity, refractive and keratometric astigmatism, and IOL rotation after 1 day, 1 month, 3 months, and long-term (mean, 13.3±5.0 months) follow-up were evaluated.ResultsAt final follow-up, 73.3% of eyes showed an uncorrected visual acuity of 20/25 or better. The postoperative keratometric value was not different from the preoperative value; mean refractive astigmatism was reduced to -0.28±0.38 diopter (D) from -1.28±0.48 D. The mean rotation of the toric IOL was 3.45±3.39 degrees at final follow-up. One eye (3.3%) exhibited IOL rotation of 10.3 degrees, the remaining eyes (96.7%) had IOL rotation of less than 10 degrees.ConclusionsEarly postoperative and long-term follow-up showed that implantation of the AcrySof toric IOL is an effective, safe, and predictable method for managing corneal astigmatism in cataract patients.