Abstract
We developed an X-ray imaging spectrometer with a high timing resolution. CMOS image sensors, widely used in optical imaging, have an excellent time resolution with considerably low readout noise. We performed experiments using a Zyla CMOS image sensor manufactured by Andor Technology pic. (UK). It employs a front illuminated(FI) scientific CMOS (sCMOS) with each pixel size of 6.5×6.5μm<sup>2</sup>. We found that nearly all X-ray events spread more than 5×5 pixels and obtained good energy resolution of 260 eV at 5.9 keV at room temperature. The spectrum shows no degradation at lower energy, which imply that there is negligibly small charge loss. We report photon counting performance for a broad range of X-rays from 3-25keV at room temperature.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.