Abstract
A HDTV camera having a direct-sensing x-ray high-gain avalanche rushing amorphous photoconductor (HARP) tube was used, for the first time, to acquire x-ray phase maps. The tube can achieve a high sensitivity as a result of the avalanche multiplication process in the HARP target. A beryllium plate, rather than a glass plate, was used as the face plate of the tube to minimize the loss of x-rays due to absorption, and a 15 µm thick HARP target was directly formed on it. In the experiment, the x-ray phase shifts produced by a rat liver were measured using synchrotron x-rays (λ = 0.0766 nm) and a triple Laue-case (LLL) x-ray interferometer. Interference patterns produced by the sample were observed with the direct-sensing x-ray HARP tube camera. A voltage of 1300 V was applied to the HARP target to give an output signal gain of two. The camera was operated in 1125 scanning-line mode, and real-time images were stored on a workstation at a rate of 30 images/s with an image format of 960 (H) × 1100 (V) pixels. A phase-map image of the sample was successfully obtained using the fringe scanning method and phase unwrapping. The observed phase shifts ranged from 50° to 200°. Trees of blood vessels in the rat liver were clearly depicted without using a contrast agent. The spatial resolution of the x-ray camera was estimated to be better than 35 µm in the vertical direction and 100 µm in the horizontal direction.