Abstract
In the last decades, interest has grown in the development of detectors that can accurately measure energy and track of low-energy (1–100 keV) charged particles. In this paper we present our effort to construct a resistive MICROMEGAS detector with a wide amplification gap (192μm), to achieve high gain with a low discharge probability, even when operating with a low-pressure gas down to 50mbar. The detector response (gain, energy resolution, relative primary charge collection efficiency) to X-rays from a 55Fe source was measured under different operating conditions of amplification and drift fields, gas pressure and gas mixture composition. The results obtained and their temporal stability demonstrate that the detector can be used for the measurement of low-energy radiation with good energy resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.