Abstract

We tested whether work of breathing in premature newborns estimated by phase angle (θ) by using respiratory inductance plethysmography is decreased during neurally-adjusted ventilatory assist (NAVA) noninvasive ventilation (NIV) versus NIV alone. NAVA NIV and NIV were applied in random order while using respiratory inductance plethysmography to measure the phase angle. Patient-ventilator asynchrony was decreased during NAVA NIV; however, the phase angle was not different between the modes. A large number of repeated assists with switches to backup were found when using NAVA NIV. Results of the analysis indicated these were due to the apnea alarm limit set during NAVA NIV. The improvement in patient-ventilator synchrony supports the hypothesis that work of breathing may be decreased with NAVA NIV; however, we were unable to demonstrate this with our study design. Short apnea time settings with NAVA NIV led to a large number of switches to backup and repeated assists during the same neural effort. (ClinicalTrials.gov registration NCT02788110.).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.