Abstract

Wear of the modular taper between head and shaft has been related to clinical failure resulting from adverse reactions to metallic debris. The problem has become pronounced in large metal-on-metal bearings, but the mechanism has not yet been fully understood. We analyzed retrieved components from five patients revised with various diagnoses. Two distinct wear patterns were observed for the head tapers. Three samples demonstrated "asymmetric" wear towards the inner end of the head taper. The other two showed "axisymmetric" radial wear (up to 65 µm) presenting the largest wear volumes (up to 20 mm(3)). Stem tapers demonstrated relatively little wear, and the fine thread on the stem taper surface was observed to be imprinted on the taper inside of the head. Our findings demonstrate that the cobalt-chrome head wears preferentially to the titanium stem taper. "asymmetric" wear suggests toggling due to the offset of the joint force vector from the taper. In contrast, samples with "axisymmetric" radial wear and a threaded imprint suggested that corrosion led to head subsidence onto the stem taper with gradual rotation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.