Abstract
Extrinsic warming of contrast media (CM) to 37 °C before angiographic procedures is performed to improve bolus kinetics and avoid potential adverse effects. Extrinsically warmed CM readily loses temperature after removal from the warming cabinet, but the extent of its cooling has not been previously investigated. To assess temperature loss of extrinsically warmed CM in tubing of traditional angiographic manifolds during simulated angiography. In total, 35 scheduled diagnostic angiographic procedures were observed in a hospital setting. Relevant time points of CM use during the procedures were recorded. The shortest, median, and longest procedures were then simulated in the experimental laboratory to measure CM temperatures at specific times at three locations along the tubing system. The angiographic procedures lasted 7.0-26.6 min (median = 11.7 min), with the total duration dependent primarily on the time from contrast being removed from the warming cabinet to the commencement of imaging. During the simulated procedures, consistent patterns of temperature loss were observed. By the last simulated angiographic run, injected CM temperature decreased by 7.4-16.4 °C, depending on procedure length. Most of the heat loss occurred in the tubing between the CM bottle and coronary control syringe. During angiographic procedures, prewarmed CM loses its temperature rapidly with the duration of exposure to ambient room temperature. If no additional measures are employed to maintain its temperature outside of the warming cabinet, extrinsic warming has limited impact on injected CM temperature.