Abstract

The EGSnrc Monte Carlo user-code CSnrc is used to calculate wall correction factors, Pwall,, for parallel-plate ionization chambers in photon and electron beams. A set of Pwall values, computed at the reference depth in water, is presented for several commonly used parallel-plate chambers. These values differ from the standard assumption of unity used by dosimetry protocols by up to 1.7% for clinical electron beams. Calculations also show that Pwall is strongly dependent on the depth of measurement and can vary by as much as 6% for a 6 MeV beam in moving from a depth of dref to a depth of R50. In photon beams, where there is limited information available regarding Pwall for parallel-plate chambers, CSnrc calculations show Pwall values of up to 2.4% at the reference depth over a range of photon energies. The Pwall values for photon beams are in good agreement with previous estimates of the wall correction but have much lower statistical uncertainties and cover a wider range of photon beam energies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.