Abstract

The EGSnrc Monte Carlo user-code CSnrc is used to calculate wall correction factors, Pwall, for thimble ionization chambers in photon and electron beams. CSnrc calculated values of Pwall give closer agreement with previous experimental results than do the values from the standard formalism used in current dosimetry protocols. A set of Pwall values, computed at the reference depth in water, is presented for several commonly used thimble chambers. These values differ from the commonly used values by up to 0.8% for megavoltage photon beams, particularly for nominal beam energies below 6 MV. The sleeve effect, which is not currently taken into account by the TG-51 dosimetry protocol, is computed to be up to 0.3% and is in some cases larger than the Pwal1 correction itself. In electron beams, where dosimetry protocols assume a wall correction of unity, CSnrc calculations show Pwall values of up to 0.6% at the reference depth, depending on the wall material. Pwall is shown to be sensitive to the depth of measurement, varying by 2.5% for a graphite-walled cylindrical Farmer-like chamber between a depth of 0.5 cm and R50 in a 6 MeV electron beam.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.