Abstract

The aim of the study was to establish the reference window settings for display of virtual monoenergetic images (VMIs) from spectral detector computed tomography when assessing hypodense liver lesions. In patients with cysts (n = 24) or metastases (n = 26), objective (HU, signal-to-noise ratio [SNR]) and subjective (overall image quality, lesion conspicuity and noise) were assessed. Furthermore, 2 readers determined optimal window center/width (C/W) for conventional images (CIs) and VMIs of 40 to 120 keV. Center/width were modeled against HUliv with and without respect to the keV level (models A and B). Attenuation and SNR were significantly higher in low-keV VMIs and improved overall image quality and lesion conspicuity (P ≤ 0.05). Model B provided valid estimations of C/W, whereas model A was slightly less accurate. The increase in attenuation and SNR on low-keV VMIs requires adjustment of C/W, and they can be estimated in dependency of HUliv using linear models. Reference values for standard display of VMIs of 40 to 120 keV are reported.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.