Abstract

ObjectivesTo evaluate the analytical performance of the TB-IGRA® assay on the VIDAS3 platform (bioMérieux) when testing a predominantly low risk population in a low incidence area. ResultsEighty-eight percent of the results were concordant between QuantiFERON®-TB Gold-Plus (QFT®-Plus, QIAGEN) and TB-IGRA®. All 12 of 99 (12.1%) discordant results were determined positive only with the TB-IGRA® assay. In 11 of 12 of these discordant cases, no explanation could be found in the medical record. Five of these discrepant results were probably caused by the use of contaminated stimulation reagents. The remaining 6 discrepant samples were also part of the reproducibility experiment and only 2 results were reproducible positive. Overall, in the reproducibility experiment 5 of 25 (20.0 %) results were not repeatable. Conclusionsthe TB-IGRA® assay seems prone to contamination. Besides, we documented a reproducibility of only 80.0% with the TB-IGRA® assay.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.