Abstract

In this paper, the effect of ground vibration on hardness measurement, Rockwell scale C hardness, Vickers scale HV1 hardness and Leeb hardness is studied. The hardness machines were placed on the vibration table. The vibration signal is single frequency sinusoidal wave, which frequency and amplitude of vibration can be controlled. The hardness value at free from vibration state is used as a reference to calculate the error of each hardness measurement at certain frequency and amplitude. Two Rockwell hardness testing machines are used to measure hardness blocks: 20, 40, and 60 HRC. Both machines give the same tendency. Significant negative errors occur around frequency 5–15 Hz. Moreover, ground vibration has more impact on the soft range of hardness than hard range. The result from this paper can be used as guideline for laboratory to control environmental vibration amplitude to be less than 0.01 m/s 2 for frequency (10 ± 5) Hz and 0.05 m/s 2 for other in Rockwell scale C hardness measurement. Effect of vibration on Vickers and Leeb hardness measurement is preliminarily studied. Results on 200 HV1 and 900 HV1 measurement show that vibration creates peak of error at frequency 20 Hz of amplitude more than 0.01 m/s 2 and the effect can be observable at all frequency for amplitude of 0.04 m/s 2. Different from Rockwell and Vickers, Leeb hardness measurement is not influenced by vibration in this experiment.