Abstract

An x-ray beam line BL-3A at the Photon Factory has been designed and constructed to have optical components of a collimating paraboloidal mirror, a sagittal-focusing double-crystal monochromator, and a focusing paraboloidal mirror. The evaluation at the (+,−,−) setting of Si (111) showed that the vertical angular divergence was reduced from 13.4 to 9.5 (arcsec) by the collimating mirror. The vertical focusing with two mirrors gave the beam size of 0.2 mm (1/50). The horizontal beam size was reduced to 1/60 by sagittal focusing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.