Abstract

The primary goal of our study was to verify the analytical performance of intact parathyroid hormone measurements conducted on the Abbott CI analyzer using the immuno-chemiluminescence method. This verification took place in the biochemistry laboratory at Mohammed VI University Hospital in Oujda. Our methodology was aligned with the recommendations outlined in the French accreditation committee (COFRAC) technical guide (GTA) 04, emphasizing the assessment of reproducibility and repeatability. Overall, the results of this evaluation were satisfactory and met the standards set by both the supplier and the French Society of Clinical Biology. This study illustrates that the biochemistry laboratory at Mohammed VI University Hospital of Oujda is equipped to provide accurate and precise results, which are crucial for effective clinical diagnosis and decision-making.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.