Abstract

Interferometric test of large convex aspheres with high accuracy is still an urgent problem. Adopting a large-aperture null corrector is usually inevitable, and it is imperative to certify the large-aperture null optics in case incorrect final shape of the aspheric mirror is obtained. Moreover, it is necessary to calibrate and remove errors of the large-aperture null lens and trace errors to surfaces that are easy to obtain high accuracy. For these purpose, this paper presents an in situ calibrated null test method. The large-aperture null corrector for convex aspheric mirror is verified and calibrated by a wisely designed small aperture certifying null whose surfaces are either flat or spherical which are easy to be fabricated, measured and assembled. A redundant test is implemented by a Zygo VeriFire Asphere interferometer for cross test. Compared with existing methods of certificating null correctors which utilize an expensive CGH or a self-aligning aspherical mirror, the presented method finally traces errors of a large-aperture null corrector to small aperture flat and spherical surfaces thus costs are expected to be saved dramatically.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.