Abstract
Vector flow mapping (VFM) is an echocardiographic approach for visualizing two-dimensional cardiac flow dynamics by estimating the radial component of flow from the Doppler velocities and wall motion velocities using mass-conservation equations. Although VFM provides two-dimensional flow in a clinically suitable, fast, and easy manner, it is only applicable to bounded regions such as ventricles. In this study, the VFM algorithm is modified so that the velocities are estimated regardless of the flow geometry. The proposed method, vascular VFM, operates on the same principle as VFM but with additional boundary conditions created by using Doppler velocities measured at a different steering angle. The method was optimized and validated using a common carotid artery phantom and particle-image-velocimetry (PIV). The results indicated that the optimal angle of an additional Doppler beam ranged from 10 to 20 degrees. Under this optimal-beam-angle condition, the standard deviation (SD) of total vascular VFM error,...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.