Abstract

BackgroundThe accuracy of vector flow mapping (VFM) was investigated in comparison to stereo particle image velocimetry (stereo-PIV) measurements using a left ventricular phantom. VFM is an echocardiographic approach to visualizing two-dimensional flow dynamics by estimating the azimuthal component of flow from the mass-conservation equation. VFM provides means of visualizing cardiac flow, but there has not been a study that compared the flow estimated by VFM to the flow data acquired by other methods.MethodsA reproducible three-dimensional cardiac blood flow was created in an optically and acoustically transparent left-ventricle phantom, that allowed color-flow mapping (CFM) data and stereo-PIV to be simultaneously acquired on the same plane. A VFM algorithm was applied to the CFM data, and the resulting VFM estimation and stereo-PIV data were compared to evaluate the accuracy of VFM.ResultsThe velocity fields acquired by VFM and stereo-PIV were in excellent agreement in terms of the principle flow features and time-course transitions of the main vortex characteristics, i.e., the overall correlation of VFM and PIV vectors was R = 0.87 (p < 0.0001). The accuracy of VFM was suggested to be influenced by both CFM signal resolution and the three-dimensional flow, which violated the algorithm’s assumption of planar flow. Statistical analysis of the vectors revealed a standard deviation of discrepancy averaging at 4.5% over the CFM velocity range for one cardiac cycle, and that value fluctuated up to 10% depending on the phase of the cardiac cycle.ConclusionsVFM provided fairly accurate two-dimensional-flow information on cardio-hemodynamics. These findings on VFM accuracy provide the basis for VFM-based diagnosis.

Highlights

  • An understanding of left ventricle (LV) flow dynamics, which is known to be multidirectional, asymmetrical, and vortical [1], will enable diagnosis of cardiac abnormalities

  • The accuracy of vector flow mapping (VFM) was suggested to be influenced by both color-flow mapping (CFM) signal resolution and the three-dimensional flow, which violated the algorithm’s assumption of planar flow

  • Statistical analysis of the vectors revealed a standard deviation of discrepancy averaging at 4.5% over the CFM velocity range for one & Tomohiko Tanaka tomohiko.tanaka.nx@hitachi.com

Read more

Summary

Introduction

An understanding of left ventricle (LV) flow dynamics, which is known to be multidirectional, asymmetrical, and vortical [1], will enable diagnosis of cardiac abnormalities. Several reliable methods of visualizing intracardiac flows [3], such as cardiac magnetic resonance (CMR), echocardiography particle image velocimetry (echo-PIV), and vector-flow mapping (VFM), have been developed. Of these methods, VFM can and non-invasively visualize 2D blood flows [4] as the method is based on conventional ultrasound scanner and does not require administration of contrast agents. VFM data is obtained on the basis of blood-flow velocity in the LV measured by color-flow mapping (CFM) and cardiac wall velocity acquired by tissue tracking. Methods A reproducible three-dimensional cardiac blood flow was created in an optically and acoustically transparent left-ventricle phantom, that allowed color-flow mapping (CFM) data and stereo-PIV to be simultaneously acquired on the same plane. A VFM algorithm was applied to the CFM data, and the resulting VFM estimation and stereo-PIV data were compared to evaluate the accuracy of VFM

Objectives
Methods
Results
Discussion
Conclusion