Abstract
Lab research might benefit from the advantages of wearable devices, such as their ease of use, to estimate pulse rate (PR) and pulse rate variability (PRV) as an equivalent for heart rate (HR) and heart rate variability. However, before implementing them in a lab context, the validity of the PR and PRV, also on ultra-short time scales (e.g., 30s), needs to be confirmed. We recorded heart activity simultaneously with an E4 wristband and an ECG device in a seated resting condition for 5 min. Our results showed that HR, RMSSD, SDNN and LF, but not HF, were validly estimated by the E4 wristband. Furthermore, the E4 wristband could validly estimate PR with recording lengths as short as 10s. RMSSD and SDNN were validly estimated using 30s or 120 s or an average of multiple short intervals (10s), while HF likely requires longer recording intervals. Based on this study, we formulated several recommendations for using the E4 wristband in a lab context.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.