Abstract

We report here validation of the Immulite 2000 Xpi cortisol immunoassay (Siemens; with kit lot numbers <550) for measurement of urine cortisol in dogs, with characterization of the precision (CV), accuracy (spiking-recovery [SR] bias), and observed total error (TEo = bias + 2CV) across the reportable range. Linearity assessed by simple linear regression was excellent. Imprecision, SR bias, and TEo increased markedly with decreasing urine cortisol concentration. Interlaboratory comparison studies determined range-based (RB) bias and average bias (AB). The 3 biases (SR, RB, and AB) and resulting TEo differed markedly. At 38.6 and 552 nmol/L (1.4 and 20 μg/dL), between-run CVs were 10% and 4.5%, respectively, and TEoRB were ~30% and 20%, respectively, similar to observations in serum in another validation study. These analytical performance parameters should be considered for urine cortisol:creatinine ratio (UCCR) result interpretation, given that, for any hypothetical errorless urine creatinine measurement, the error % on UCCR mirrors the error % on urine cortisol. Importantly, there is no commonly used interpretation threshold for UCCR, given that UCCR varies greatly depending on measurement methods and threshold computation. To date, there is no manufacturer-provided quality control material (QCM) with target values for urine cortisol with an Immulite; for Liquicheck QCM (Bio-Rad), between-run imprecision was ~5% for both QCM levels. Acceptable QC rules are heavily dependent on the desired total allowable error (TEa) for the QCM system, itself limited by the desired clinical TEa.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call