Abstract
Few studies have been conducted on blood pressure monitors and their use at high altitude. This study is the first to evaluate an automated blood pressure device at high altitude following a standard validation protocol. The Omron HEM-7201 upper arm automatic blood pressure monitor was tested for accuracy in Lhasa, Tibet, China (3650 m above sea level) according to the European Society of Hypertension International Protocol revision 2010 (ESH-IP2). Thirty-three participants received 9–10 sequential blood pressure measurements alternating from a mercury sphygmomanometer and the device. The mean device-observer measurement difference was 1.0 ± 5.9 mmHg for systolic blood pressure (SBP) and −3.1 ± 4.6 mmHg for diastolic blood pressure (DBP). Of the 99 measurement pairs analyzed, 72, 90, and 97 device readings were within 5, 10, and 15 mmHg, respectively, of the observer measurements for SBP, and 68, 92, and 99 readings for DBP. The number of participants with at least two out of three measurements within 5mmHg was 27 for SBP and 25 for DBP. Three participants had no measurements within 5 mmHg for SBP or DBP. The Omron HEM-7201 passes the ESH-IP2 validation criteria and can therefore be recommended for use in adults in this setting.
Published Version (
Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have