Abstract

The purpose of this study was to determine the performance characteristics of the Cozart ® microplate EIA for detecting opiates in oral fluid from patients in a drug misuse treatment program. Oral fluid samples were collected using the Cozart ® RapiScan Collection System from 216 donors who were receiving treatment for their addiction and were monitored for drug abuse. A further 40 oral fluid samples were collected from volunteer donors who were not drug users. The samples were analyzed in the laboratory by using the Cozart ® microplate EIA for opiates and confirmed using gas chromatography–mass spectrometry (GC–MS). The samples were stored frozen until analysis by GC–MS. The intra-assay precision for the Cozart ® microplate oral fluid EIA for opiates over 40 assays was 0.43% to 9.13% CV (within assay) and 2.9% to 9.1% CV (within day). A total of 109 samples were positive for various opiates. The Cozart ® microplate EIA for opiates in oral fluid, using a cut-off of 30 ng/ml morphine equivalents in neat oral fluid, had a sensitivity of 99.1 ± 2.1% and a specificity of 94.4 ± 2.2% versus GC–MS. A series of potential adulterants of oral fluid were evaluated and shown not to alter the outcome of the test result.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.