Abstract
To validate the Andon KD-5917 automatic upper arm blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. Sequential same-left-arm measurements of systolic blood pressure (SBP) and diastolic blood pressure (DBP) were obtained in 33 participants using the mercury sphygmomanometer and the test device. According to the validation protocol, 99 pairs of test device and reference blood pressure measurements (three pairs for each of the 33 participants) were obtained in the study. The device produced 73, 98, and 99 measurements within 5, 10, and 15 mmHg for SBP and 86, 98, and 99 for DBP, respectively. The mean ± SD device-observer difference was 3.07 ± 3.68 mmHg for SBP and -0.89 ± 3.72 mmHg for DBP. The number of patients with two or three of the device-observer difference within 5 mmHg was 26 for SBP and 29 for DBP, and no patient had a device-observer difference within 5 mmHg. The Andon KD-5917 automatic upper arm blood pressure monitor can be recommended for clinical use and self-measurement in an adult population on the basis of the European Society of Hypertension International Protocol revision 2010.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.