Abstract

The aim of this study was to validate the iHealth BP5 wireless upper arm blood pressure (BP) monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010. The ESH-IP revision 2010 for validation of BP measuring devices in adults was followed precisely. A total of 99 pairs of test device and reference BP measurements (three pairs for each of the 33 participants) were obtained in the study. The device produced 71, 89, and 97 measurements within 5, 10, and 15 mmHg for systolic blood pressure (SBP) and 73, 90, and 99 mmHg for diastolic blood pressure (DBP), respectively. The mean ± SD device-observer difference was -1.21 ± 5.87 mmHg for SBP and -1.04 ± 5.28 mmHg for DBP. The number of participants with two or three device-observer differences within 5 mmHg was 25 for SBP and 28 for DBP. In addition, three participants had no device-observer difference within 5 mmHg for SBP and none of the participants had the same for DBP. According to the validation results on the basis of the ESH-IP revision 2010, the iHealth BP5 wireless upper arm BP monitor can be recommended for self/home measurement in an adult population.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.