Abstract

The goal of this work was to validate the method of the k 0 standardization in neutron activation analysis at the laboratory of neutron activation analysis in Řež using the Kayzero for Windows software package. For verification of the precision and accuracy of the method two types of reference materials were used. One was a set of synthetic materials aimed at checking the implementation of k 0 standardization. The others were matrix standard reference materials of the US National Institute of Standards and Technology comprising several types of matrices. In general, good agreement was obtained between results of this work and values in certificates of the individual reference materials, thus proving the accuracy of our results and successful implementation of the method for analysis of real samples. In addition, detection limits for more than 50 elements were evaluated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.