Abstract

A simple, feasible, isocratic elution, and stable reversed-phase high performance liquid chromatography method was established and verified. The chromatographic conditions are as follows: EF-C18H, 4.6 × 250 mm, 5 μm column; column temperature 30 °C; for the mobile phase 27.2 g of KH2PO4 and 8.5 g of tetrabutylammonium hydrogen sulfate were taken, 2500 mL of water was added to dissolve, and the pH was adjusted to 6.7 with phosphoric acid:methanol solution with a ratio of 84:16 (V:V). The flow rate was 1.0 mL/min; the injection volume was 10 μL; and the wavelength was 262 nm. According to the current ICH guidelines, the developed method was verified, and the system suitability, specificity, LOD, LOQ, linearity, range, accuracy, repeatability, durability, and solution stability of the proposed method were verified. The validation results demonstrated that the LOQ for the method was 0.05% and the LOD was 0.02%. The content was detected within the concentration range of 300 to 900 μg/mL. The relationship between concentration and measurement was linear, with an r2 of >0.999. The concentration of impurities ranged from 0.3 to 4.5 μg/mL. A good linear correlation was observed within the range of g/mL, with a coefficient of determination r2 greater than 0.999. The accuracy and repeatability met the specified criteria.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.