Abstract

Urinary trans, trans-muconic acid ( t, t-MA), a biomarker of benzene exposure, is usually determined by HPLC methods with detection by either UV or, more recently, electrospray tandem mass spectrometry. However, not all these methods have been fully validated for quantitative analysis. This paper presents an HPLC/MS/MS method for reliable quantitative determination of t, t-MA that uses a commercial deuterium-labeled isotope as internal standard; the matrix effect has been evaluated and LOD is 0.22 μg/L. We used this method to test 200 urine samples, 175 of them collected at end-of-shift from workers in an oil refinery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.