Abstract
This study validated an automated method of finite element modelling of bone from CT scan data. After a fresh-frozen cadaveric femur was modelled, strain gauges were attached to the bone at 11 locations and the femur was mechanically tested by applying a load to the femoral head. Linear regression analysis was used to correlate the strains predicted by the model with the experimentally measured strains. The regression results were significant ( P < 0.001), indicating that the strain calculated by the FE model is a valid predictor of the measured strain. Verification of the surface strains also supports the validity of the strains and stresses predicted inside the bone. The present study provides a strong rationale for use of this modelling method as a research tool and in possible clinical applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.