Abstract

The method for the determination of caprylic acid and sodium caprylate from biological products was systematically validated using NEFA-C kit. The results obtained demonstrated that the kit method was simple, rapid, reliable, sensitive, reproducible and cost effective in comparison to the current methods i.e. colorimetric, High Performance Liquid Chromatography (HPLC) and Gas Chromatography (GC) methods. The assay exhibited excellent linearity, accuracy, precision and robustness. Mean recoveries ranged between 95 and 101.3% ( n = 6). The proposed method was linear over the concentration range of 0.05–10 mM of caprylate with values of coefficient of regression being >0.99. Method showed sensitivity of 0.05 mM (7.21 μg/ml for caprylic acid and 8.31 μg/ml for sodium caprylate). The % Relative standard Deviation (%RSD) for intra and interprecision studies was less than 5%. In conclusion the validated method was successfully used in monitoring of processed bulk and final products generated during production of biological products thus laying emphasis on strict control of release criteria for biological products fractionated using caprylic acid.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.