Abstract
High throughput testing of clinically representative Pt electrodes requires an inexpensive, efficient method of production. The aim of this study was to develop a facile platinum (Pt) model electrode (PME) and assess its production process, stability, and reproducibility. In this study a new model electrode was developed using representative substrates and dimensions as state-of-the-art electrode arrays used for neural stimulation. It was found that the PME is a highly reproducible robust system with similar electrochemical performance but with lower variability than other neural prosthetic arrays.Clinical Relevance- As an estimate these novel model electrodes cost 300 times less than a cochlear implant, can be manufactured in a tenth of the time and with a less than 10% failure rate. It is expected that model electrodes with low variability of electrical properties will significantly improve preclinical validation testing of electrochemical stimulation, surface modifications, and coatings.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.