Abstract
Quantitation of ADAMTS13 activity, functional inhibitors, and autoantibodies is crucial in diagnosis and management of thrombotic thrombocytopenic purpura. We compared and optimized commercial assay kits and validated a testing panel. Citrated plasma specimens from healthy volunteers and residual samples submitted for clinical testing were used in the study. Commercially available ADAMTS13 activity assays including ACTIFLUOR(™) ADAMTS13 (Sekisui Diagnostics, Stamford, CT, USA), LIFECODES ATS-13 (Gen-Probe Inc., San Diego, CA, USA), and TECHNOZYM(®) ADAMTS-13 (Technoclone, Vienna, Austria) were evaluated. Functional inhibitor assays were performed using internally developed mixing protocols. Two autoantibody assays were also evaluated: IMUBIND(®) (Sekisui Diagnostics) and TECHNOZYM(®) ADAMTS-13 INH ELISA kits (Technoclone). A laboratory-developed assay using ACTIFLUOR(™) reagents showed best agreement with the reference method, and full validation showed a reportable range of 5% (LLOQ) to 114% with a reference interval of ≥68%. Both intra- and interassay coefficients of variation were <10%. Inhibitor assays performed with the kits showed 95% overall agreement with the reference method. A modification of the TECHNOZYM(®) autoantibody assay showed 85% overall agreement with the reference method with imprecision approximately 20%. ADAMTS13 activity and inhibitor tests using ACTIFLUOR(™) reagents and modified TECHNOZYM(®) autoantibody ELISA showed superior performance compared to the other kits for clinical use in this study.