Abstract

Aim: Allograft rejection remains a major cause of graft failure in kidney transplantation. Here the authors report the validation of a non-invasive molecular diagnostic assay, AlloMap Kidney, using peripheral blood. Methods: The AlloMap Kidney test is a gene expression profile utilizing the RNA-seq platform to measure immune quiescence in kidney transplant patients. Results/Conclusions: Analytical validation showed robust performance characteristics with an accuracy correlation coefficient of 0.997 and a precision coefficient of variation of 0.049 across testing. Clinical validation from the prospective, multi-center studies of 235 samples (66 rejection and 169 quiescence specimens) demonstrated the sensitivity of 70% and specificity of 66% for allograft rejection, while the negative predictive value was 95% to discriminate rejection from quiescence at 10% prevalence of rejection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.