Abstract

The present study describes new, simple, rapid and novel spectrophotometric methods for simultaneous estimation of Tramadol Hydrochloride and Ketorolac Tromethamine combination was analysed by simultaneous equation method (Method B) and First order derivative spectroscopic method (Method C). Method A involved measurement of absorbance at two wavelengths, 271 nm and 246 nm, λmax of tromethamine respectively. For method B, wavelengths 331 nm (for determination of Ketorolac tromethamine) and 271 nm were used. For method C, wavelengths 339.2 nm (Zero crossing of Tramadol hydrochloride) and 229.4 nm (Zero crossing of Ketorolac tromethamine) were used for determination of Ketorolac tromethamine and Tramadol hydrochloride respectively. Beer's law was obeyed in concentration range of 5–50 μg/ml and 2–20 μg/ml for Tramadol hydrochloride and Ketorolac tromethamine respectively by all the methods. Method C was found to be more simple and sensitive. The proposed methods are pharmaceutical formulations due to their specificity, rapidity, simplicity and accuracy. These methods validated for linearity, accuracy and precision as per ICH guidelines.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.